ORION: Test Diffractometer

orion-cradle.jpg

ORION is an in-house two axis diffractometer that can be used to check and align samples. The sample sits on an easily accessible sample table and can be aligned using the two gonios. The instrument has a single He3-tube detector and the collimation can be easily changed to allow very fine tuning of alignments.

To use ORION please speak to your local contact or the instrument responsible.

New commands
initaux:initiate auxiliary UBAn initial UB matrix to get you quickly underway with your alignment
  initaux h1 k1 l1 h2 k2 l2
addauxref:add auxiliary reflectionsAdd auxiliary reflections to reflist, these are assumed to be in the scattering plane and hence the phi, omega and 2theta angles are calculated.
  addauxref h k l
home:drives motors to rest positionOmega, Chi and Phi are driven to zero, 2theta is driven to 90. The instrument should be driven to zero once an alignment is complete.
  home
mv:relative movement of a motorMove a motor relative to the current position, rather than to an absolute value of omega, chi, phi or 2theta.
  mv motor degrees

Instrument phone numbers

ORION area:
+41 56 310 37 82

ORION cabin:
+41 56 310 31 89