Calls for Proposals

The PSI User Office invites user proposals for the next user run at SwissFEL

  • We call for proposals for the Alvra, Bernina, and Cristallina instrument at the Aramis hard X-ray line and for the Maloja and Furka instrument at the Athos soft X-ray line. Please take particular note of the restricted configurations for this run described below and on the respective instrument tabs.
  • Newly, the Cristallina-Q instrument ultralow-T vectormagnet is offered to the users’ community.  It is designed for X-ray diffraction of quantum condensed-matter systems at sub-Kelvin temperatures and in magnetic fields up to 5 Tesla. 
  • SwissFEL is being continuously developed but the proposals must be based on the parameters outlined below on this page.
  • After the submission deadline the proposals are evaluated in terms of safety and technical feasibility. Then they are ranked in terms of scientific criteria by the international SwissFEL Proposal Review Committee (PRC). More information about the evaluation procedure is published on Evaluation. The result of this rating is the basis for the beamtime assessment made by SwissFEL.
  • The main proposers are informed by email about the result of the ranking and beamtime assessment.
  • The annual calendar for the proposal evaluation is as seen below.
  • Proposals can be submitted to the PSI User Office.

For the current call, we can only offer restricted instrument capabilities. 

In early July a small fire has damaged a modulator of the SwissFEL injector and our subsequent machine studies have shown that we will not be able to offer beam to users with the quality and reliability that we expect from SwissFEL until all repairs have been finished. As a result, we need to reschedule a significant number of experiments into the first semester of 2025 and thus the number of available beamtime slots will be reduced. 

In order to serve the overall users’ community well and to operate as efficiently as possible during the following run, we will offer the following capabilities for this call:

  • ARAMIS-Alvra: Prime endstation with liquid jet spectroscopy setup
  • ARAMIS-Bernina: High-field THz cryo chamber setup or Surface Diffractometer setup
  • ARAMIS-Cristallina-MX: Fixed target SFX setup
  • ARAMIS-Cristallina-Q: Ultralow-T Vectormagnet setup
  • ATHOS-Maloja: Time-resolved XPS, ion momentum spectroscopy or single-particle imaging or time-resolved XAS
  • ATHOS-Furka: RIXS setup, REXS and XAS setup 

Schedule for Calls

We will open a call for proposals for SwissFEL in February 2025. 

SwissFEL call schedule
Experimental Period01.08.2025 - 31.12.2025  
Callapprox. 10.02.2025  
Submission deadline15.03.2025  
Start period01.08.2025  
End period31.12.2025  
EVALUATION
 
23. - 24.04.2025  

Aramis beam parameters

Regular SASE:

  • photon energy: 1.8 - 12.7 keV
  • polarization: linear horizontal
  • typical relative bandwidth (cumulated): 0.25% fwhm 
  • pulse energy: typically up to 1000 µJ
  • repetition rate: single shot - 100Hz  (accelerator at 100 Hz, rate reduction via fast shutter) 
  • pulse duration: 40-70 fs fwhm
  • X-ray pump-laser arrival-time jitter:  < 150 fs fwhm (for time tool options see instrument sections)

Advanced machine modes with lower operation experience are available on a best effort basis. These include:

  • Large bandwidth mode: relative bandwidth up to 2%
  • Short pulse mode: pulse length 10-40 fs fwhm, pulse energy scales with pulse length
  • Ultrashort pulse mode: pulse length < 1 fs fwhm, typical pulse energy 5 – 10 uJ

In case you plan to submit a proposal which requires an advanced mode, we advice to consult with the endstation contacts beforehand.

 

For the new run the Alvra, Bernina and Cristallina endstations are available with the following parameters:

The Alvra instrument specializes in ultrafast dynamics in chemical and biological systems, especially in solutions, liquids, or crystals in viscous media. Alvra is equipped for X-ray spectroscopy (absorption and emission), liquid scattering, and serial femtosecond crystallography measurements, as well as flexible user-provided setups. For the current call, the Alvra Prime instrument (vacuum and He environment) is available for liquid jet spectroscopy setups. The details are described below.

Alvra Prime
Photon energy range2 keV – 12.4 keV (fully commissioned over the full energy range)  
Beam profile
  • Measured focus <10 x 10 µm2 (fwhw) below 4.5 keV.
  • Measured focus  < 5 x 5 µm2 (fwhm) for energies > 6 keV
  • 1 x 1 µm2 (fwhm) achieved for 12 keV
  • Focus can be adjusted to meet experimental requirements
  • Unfocused beam: ~1 x 1 mm2 (fwhm) energy dependent
  
BandwidthMonochromatic (Si(111), InSb(111), Si(311)) and pink beam (0.25% of fundamental); larger bandwidths of up to 2% are also possible (photon energy dependent)  
EnvironmentVacuum (down to 5x10-4 mbar ) up to atmospheric pressure (He or N2)  
Sample delivery

Liquid jet:

  • Thin flat jets (5-30 µm) with HPLC pump
  • Flat jet (100, 200, and 300 µm) with peristaltic pump
  • Round jet (25-200 µm) with HPLC pump
  
Detectors and Spectrometers
  • 2 x 2 crystal von Hamos dispersive X-ray emission spectrometer (1-12.4 keV)
  • Diodes for integrated x-ray absorption measurements
  
Alvra experimental laser infrastructure
Fundamental800 nm, 35 fs (fwhm), 10 mJ (Ti:Sapphire)  
Harmonics800/400/266 nm branch available in parallel to OPA  
OPA conversion240 nm – 2.5 µm  
Pulse energy at the sample position

Measured OPA pulse energies at the sample location vary with wavelength, ranging between 5 to 100 µJ. For specific pump wavelengths please inquire. Pulse durations are expected to be approximately 75 fs fwhm.

Harmonics branch allows for higher pulse energies at 800/400/266 nm with shorter pulse durations achievable. For current status please inquire.

  
Focus50 x 50 – 500 x 500 µm2 (fwhm)  

 

General information about the Alvra endstations can be found at: Alvra
For questions and further information about Alvra contact: Dr. Camila Bacellar

The Bernina Instrument is specialized on studying condensed matter systems by selective light excitation and selective X-ray probes. It is equipped with flexible but precise positioning hardware for diffraction on solid state samples, but also larger user-supplied hardware. The instrument can interchange two endstations, which can be configured for different sample and detector configurations. Bernina features versatile pump laser excitation from UV to single cycle THz pulses.

For the call of 2025A experiments, the flexible Bernina endstation platform will be limited to the following standardized configurations:

  • Six circle kappa surface diffractometer, equipped with area detector on 2-circle detector arm (XRD). Sample in air or N2-cryostream atmosphere (80-500K stream temperature). Standard pump laser wavelengths from UV to near Ir. An optional polarisation analyzer can be used upon early request.
  • Vacuum chamber setup for low sample temperature (down to ~5K), high electric field THz pulse excitation, and tender X-ray range experiments. It is advised to explore the compatibility of proposed experiments with the chamber geometry before proposal submission. For details about the chamber please refer to https://doi.org/10.1088/1361-648X/ac08b5

Experiments requiring a custom assembly using the flexible platform are encouraged to start communication with the Instrument staff and potentially submit a proposal for early evaluation of technical and scientific feasibility of their idea, respectively. 

New developments

  • Visible short pulse pump (500 – 750 nm; ≤ 25 fs fwhm) by Non-Collinear Optical Parametric Amplifier (NOPA).
  • Load lock for High-field THz cryo chamber for quick sample exchange and transfer from inert atmosphere (under commissioning, usable upon request).

For high time resolution experiments, a single pulse timing diagnostics is used for time-delay feedback and time delay measurement close to pulse length limit. The diagnostics works reliably at Si-111 monochromatic and pink FEL beam.

Bernina
Photon energy range 2 keV – 13.0 keV
Beam profile Focused down to 2x2 µm2 (fwhm, measured) to unfocused 1000x1000 µm2 (fwhm, photon energy dependent).
Bandwidth Monochromatic (Si(111) routinely used, InSb(111), Si(311) and pink beam (~0.2% of fundamental, transmissive single FEL pulse spectrometer available), special modes like broadband SASE operation possible.
Pulse length Standard SASE pulse length ~50 fs (fwhm), short pulse options at cost of pulse energy down to ~20 fs.
Environment He or ambient atmosphere, platform for user-supplied chambers, N2 and He based cryostream coolers down to ~80 K. Vacuum chamber available for low sample Temperature (<5 K), high field THz excitation, and tender X-ray range.
Sample systems

Solids: single crystals, powders, amorphous systems. Liquid/Gas only with user supplied equipment.

Detectors and Spectrometer
  • 1.5 M Jungfrau detector
  • 16 M Jungfrau detector
  • Diodes or single element (0.5 M) Jungfrau detectors for multi- purpose applications.
  • Resonant Inelastic X-ray spectrometer with spherical diced Si(844) analyzer crystals
  • Polarisation analyzer
  • Limited X-ray emission spectrometer options upon request
Bernina optical pump laser
Source type Wavelength range Pulse energy / max. Field Pulse length Comments
Primary pump source    800 nm 20 mJ

35 fs
or 100 fs

2nd and 3rd Harmonic (400 nm and 266 nm), as well as ~10 fs compressed fundamental 800 nm available upon request.
OPA
240 – 480 nm      10 – 100 uJ   ≤ 50 fs (fwhm) Pulse energy variation includes typical losses between source and sample.
 
480 – 780 nm 150 – 1000 uJ ≤ 50 fs (fwhm)
780 – 1160 nm 10 – 100 uJ ≤ 50 fs (fwhm)
1.16 – 1.58 µm

≤ 50 fs

1.58 – 2.4 um 500 – 1000 uJ ≤ 50 fs
NOPA 500 – 750 nm 10 – 70uJ
 
≤ 25 fs  
OPA DFG

4.4 – 20 um  (70 15 THz)

1 – 20 µJ
 
≤ 300 fs  
Opt. rect. 2 – 5 THz single cycle >500 kV/cm  
LiNbO3 based

0.5 – 2 THz single cycle

>500 kV/cm  

General information can be found at: Aramis Bernina Experimental Endstations
For questions and further information about Bernina, please contact: Dr. Henrik Lemke

The first Cristallina-MX experimental station is called SwissMX (Serial WIth Solid-Support MX), a fixed-target setup designed for high-throughput SFX and SFX pump-probe. SwissMX has been commissioned with an in-air sample environment since May 2022 and is currently available for users.

The SwissMX has currently only been commissioned to accept two styles of fixed-target: the PSI MISP (MIcro-Structured Polymer) and the MPI SOS (Sheet-On-Sheet) chip (Doak et al., 2018). The MISP-chip is composed of an array of apertures in defined locations in a similar style to the Oxford (Horrell et al., 2021) and HARE chips (Mehrabi et al., 2020). The MISP-chips can be fabricated in either a clear or an opaque polymer, with the only latter being compatible with pump-probe measurements. The SOS chip by comparison is made by carefully sandwiching a sample between two sheets of polymer film and are ideal for samples of limited volume in viscous media such as LCP. Liquid-crystal samples will need to be embedded in a moderately viscous medium to be used effectively. Please indicate in the DUO which chip or both you would like to use.

The SwissMX is capable of both SFX and nanosecond resolution SFX pump-probe measurements. An EKSPLA nanosecond OPO can be used for pump-probe measurements and has been commissioned using the opaque MISP chips. Pump-probe is only possible using these chips.

Our standard SFX beamtime configurations are: X-ray photon energy of 12 keV with standard bandwidth (0.15 % SASE), nano-second laser excitation between 410 nm - 700 nm (if required), Jungfrau 8M detector and either MISP or SOS chips. Max beamtime length will be 2 days (6 shifts). Please indicate if you would only like 0.5/1 shifts for non-pump-probe SFX measurements.

SwissMX
Photon energy range
  • 10 – 12.4 keV
  • SwissMX is currently only commissioned for in air measurements
  • Therefore, photon energy practically limited to >10 keV
  • The beam is caught by scatter guards before and after the sample position, giving a 15 mm of air window
  
Beam profile
  • Smallest measured beam size to 1.5x1.5 µm2 (FWHM) at 12 keV
  • Larger beam profiles have been used but not fully commissioned – please contact the beamline scientist if you have questions
  
Bandwidth
  • Pink Beam with 0.25% of fundamental is standard; larger bandwidths of up to 2% are also possible (photon energy dependent)
  
Environment
  • SwissMX has only currently been commissioned for in air measurements
  
Fixed-targets

PSI MISP-chip

  • Small – 6,000 apertures
  • Large – 26,000 apertures
  • Aperture pitch = 120 µm (distance between adjacent apertures)
  • Typically sealed in two layers of film - 6 µm Mylar standard but thinner films are also possible.
  • Suitable for SFX and SFX pump-probe

MPI SOS chip

  • No apertures – crystals are embedded randomly in media and sandwiched between two pieces of film - 6 µm Mylar standard but thinner films are also possible.
  • User defined shot-spacing
  • 25x25 µm spacing commissioned with no radiation damage at 12 keV
  • Possible 250,000 images from 12.5x12.5 mm2
  • Not suitable for SFX pump-probe
  
Detectors
  • 8 Mpixel JUNGFRAU (125 mm radius) on an 180 mm translation stage
  • Smallest sample detector distance = 111 mm
  • Please give a reasonable estimate of your desired resolution so the detector geometry can be properly determined prior to beamtime. See Cristallina-MX for a rough guide.
  
Optical pump laser
  • EKSPLA NT230-100-SH/SF-ATTN/FC ns OPO
  • Pulse width = 3.2 ns at 450 nm
  • There are two coupling strategies:
    • on-axis to X-ray beam - only visible wavelengths
    • through-space coupling - principally for UV and IR
  • Pulse energies are limited in the fibre-coupling to 2 µJ
  • Commissioned wavelength from 350-700 nm
  • Please contact beamline scientist if you wish to move outside of the visible spectrum
  • Available time delays = 10 ns to 50 ms
  

General information can be found at the Cristallina-MX project page.
For questions and further information about Cristallina-MX, please contact: Dr. John Beale

The Cristallina-Q team has recently commissioned an experimental station named Ultralow-T Vectormagnet that enables resonant and non-resonant X-ray diffraction in high-magnetic fields and sub-Kelvin temperatures. It includes a dedicated cryostat with beryllium and mylar windows for the incident and the diffracted beams in the horizontal diffraction plane. The cryostat is installed on a heavy-load diffractometer to orient the sample with respect to the vertical axis (θ), as well as to position the detector at the Bragg angle (2θ) and at a suitable distance from the sample. The beam size on the sample can be varied by means of bendable KB focusing mirrors. The station can be placed at custom distances from the last KB mirror.

The ultralow-T sample environment requires mounting the sample and closing the cryostat at least 1 week before the beamtime start, to allow for cooling to the base temperature and offline tests. Depending on the requested measurement temperature and the sample properties, reduced repetition rates and/or high attenuation have to be used to ensure recovery of sub-Kelvin temperatures between subsequent X-ray pulses. Because of the latter, it is advantageous, although not mandatory, to operate the ARAMIS beamline in the ultrashort pulse mode such that that thermal and electronic modification of the sample can be outrun.

In this first user call, the capabilities of the Ultralow-T Vectormagnet are restricted in terms of the maximal θ range, due to mechanical limitations, and to higher photon energies to cope with in-air propagation and due to absorption of the X-ray windows.

Cristallina-Q Instruments
Photon energy range
  • ~5 keV – 12.7 keV
  
Repetition rate
  • 0.1 - 100 Hz;  the maximal rate is given by SwissFEL accelerator operation settings (normally 100 or 50 Hz), and can be reduced using the fast shutter.
  • Pulse on demand or custom pulse pattern possible via the fast shutter.
  
Beam profile on the sample
  • From focused down to < 2 x 2 µm2 (fwhm, measured) to unfocused 1 x 1 mm2 (fwhm, energy dependent).
  • Small beam size on the sample results in larger local sample heating.
  
Bandwidth
  • Pink beam (no monochromator) – 0.25% bandwidth.
  • Single-shot spectra are recorded with upstream transmissive spectrometer.
  
Pulse length & energy
  • Regular SASE mode: ~50 fs (fwhm), ~1 mJ/pulse.
  • Short pulse mode: ~10 fs (fwhm), ~200 µJ/pulse.
  • Ultra-short pulse mode: ~0.5 - 1 fs (fwhm), 5 - 10 µJ/pulse.
  
Temperature
  • 0.1 K - 2 K sample temperature (strongly linked to the beam intensity and the repetition rate, as well as sample properties).
  • Indicative sample temperatures at 1 Hz repetition rate: <100 mK at 0.1 uJ/pulse, <200 mK at 2 uJ/pulse, 500 mK at 150 uJ/pulse.
  
Magnetic field
  • 'Vertical mode': up to 5.2 T vertical field with up to 0.26 T horizontal field (2.5° tilt towards the horizontal plane in any direction).
  • 'Isotropic mode': up to 0.6 T field in any direction (full sphere).
  
Scattering geometry
  • Full access to horizontal scattering plane, as well as ±10° vertical scattering. 
  • In-situ sample rotation around the vertical axis (θ) restricted to Δθ = 20°.
  
Detectors
  • 1.5M Jungfrau detector.
  • Sample-detector distance: 250 - 700 mm with detector on the 2θ arm.
  

General information about the Cristallina-Q experimental station can be found at: Cristallina-Q.

Because of the novelty of this experimental station, the inherent challenges due to the ultralow-T environment, and operational limitations, proposers must contact the Cristallina-Q team well ahead of submission.

Athos beam parameters

Standard SASE:

  • Photon energies and pulse energies for circular polarization (linear polarization  30-40% less) : 
    • 350 - 1000eV with more than 2000μJ pulse energy
    • 1000 - 1300eV with more than 1000μJ pulse energy
    • 1300 - 1600eV with more than 250μJ pulse energy
  • Typical bandwidth ≤ 1 %
  • Repetition rate 100Hz
  • Standard X-ray pulse duration ≤ 100fs FWHM
  • Short X-ray pulses with tuneable pulse duration down to factor of x5 relative to standard SASE mode. Pulse energy directly proportional to pulse duration
  • Polarization adjustable: Circular +/- (delivers highest pulse energy), linear horizontal, linear vertical
  • Energy scan: feasible over +/- 10 % of fundamental energy

Advanced machine modes with lower operation experience are available on a best effort basis. These include:

  • Two colour X-ray mode: Two X-ray pulses with independently tuneable photon energies up to 1000eV, maximum delay between the pulses is 500fs and minimum delay is -50fs.

Please note, that advanced modes will result in overall reduced machine performance, like lower repetition rate and reduced pulse energy as compared to standard SASE operation. In case you plan to submit a proposal which requires an advanced mode, we advice to consult with the endstation contacts beforehand.

The Maloja instrument specializes on studying ultrafast processes in atomic, molecular, non-linear and chemical sciences. The flexible setup allows for a variety of spectroscopy and imaging approaches. For the current call only the following standard setups and parameters will be available:

SetupConfiguration
Time-resolved XPS
  • Hemispherical electron analyzer: Specs Phoibos 150 EP
  • Gas needle or oven for sample delivery
  • Grating spectrometer after interaction point (Scienta XES350) for single shot spectral diagnostics in pink beam mode
  • Ion time-of-flight spectrometer upstream of interaction point used for Io for monochromatic X-rays
Time-resolved XAS
  • Grating spectrometer after interaction point (Scienta XES350) in transmission geometry
  • Transmissive spectral characterization with electron spectrometer prior to sample (Specs Phoibos 150 EP)
  • Gas cell or solid sample support
Ion momentum spectroscopy
  • COLTRIMS spectrometer with short or long ion side
  • Hexagonal delay line anode for ion coincidence measurements
  • Quadratic delay line anode for ion recoil spectroscopy
  • Supersonic gas jet (Even-Lavie valve, room temperature to 100C heatable)
  • Setup is still under commissioning, please consult with Maloja contact for details
Single-particle imaging
  • 4M Jungfrau forward scattering detector, in standard configuration 34cm detector to sample distance
  • Pulsed supersonic gas jet (room temperature) or aerodynamic lens stack injectors, please consult with Maloja contact in case you plan to use the aerosol injector
Beam profile                                  
  • Tightest focus reaching 5um x 5um (FWHM)
  • Unfocused beam 3-7mm, depending on photon energy
  • Focus adjustable with bendable KB optics to meet experimental requirements                                                      
Optical laser
  • Fundamental wavelength: 800nm
  • Pulse energy 10mJ
  • Pulse duration 35fs
  • Harmonics of the fundamental at 400nm and 266nm are available
  • TOPAS in the UV and visible regime available

Natural jitter between X-ray and optical pulse ~150fs (FWHM), which can be improved with an arrival time monitor.

Restrictions for hazardous samples apply, please consult with the Maloja contact before submission in case you plan to use such substances.

General information about the Maloja endstations can be found at: Maloja
For questions and further information about Maloja contact: Dr. Kirsten Schnorr

The Furka experimental endstation, located at the soft x-ray Athos beamline of the SwissFEL, is dedicated to the study of quantum materials using time-resolved Resonant Inelastic and Elastic X-ray Scattering (tr-RIXS and tr-REXS) as well as X-Ray Absorption (tr-XAS) spectroscopy.

The endstation is equipped with at 4-circles UHV diffractometer (Tmin=25 K), a set of x-ray detectors (APDs) rotating on two independent circles around the sample and 6 meter long RIXS spectrometer (200 meV min energy resolution).  Sample cleaving in the load lock (room temperature, vacuum 10-7 mbar) is available. The THz radiation is focused using a parabolic mirror mounted on a motorized manipulator (5 DOF) located inside the vacuum chamber.

Natural jitter between X-ray and optical pulses ~150fs (FWHM). For higher time resolution experiments, a single pulse timing diagnostics, for jitter and drift corrections, can be used (please contact instrument responsible for more info).

Furka
Setup Configuration
Time resolved XAS and REXS
  • Avalanche photodiode for FY detection (10x10 mm – 400 mm distance to sample position)
  • 2x10 mm photodiode for diffraction detection (400 mm distance to sample position)
  • Variable x ray polarization available (C+, C-, LV, LH)
Time resolved RIXS
  • sVLS Grating spectrometer
  • Energy resolution of  ≥ 200 meV for hv=500-1000 eV
  • Scattering angle 45-140 degrees. Angle variation under UHV conditions
  • CMOS detector 4MPixel (1 MPix @ 100 Hz acquisition), sensor size 1x1 inch2
Furka
Beam profile                                  
  • Tightest focus 10um x 10um (FWHM)
  • Unfocused beam 3-5mm, depending on photon energy
  • Focus adjustable with bendable KB optics to meet experimental requirements
  • Typical RIXS experiments focus 10 x (300-500) um
Enviroment
  • Vacuum baseline 1x10-8 mbar. For vacuum <1x10-8 mbar please contact instrument responsible 
Furka optical pump laser
Primary pump source    800 nm, 35 fs / 100 fs FWHM, 20 mJ (Ti:Sapphire)
Secondary pump sources with
35 fs 800 nm pulses
Wavelength range Pulse energy / max. Field Pulse length Comments
400 nm      200 uJ
(measured)  
~ 50 fs (fwhm)
266 nm 30 uJ
(measured)  
~ 50 fs (fwhm)
Secondary pump sources with
100 fs 800 nm pulses
240 – 780 nm   10 – 350 µJ ~ 100 fs (fwhm) Pulse energy depends highly on wavelength, for more information please contact the instrument responsible.
1.2 – 2.5 µm 200 – 500 µJ ~ 100 fs (fwhm)
~1 THz single cycle Up to 300 kV/cm
(measured)
THz generation with nonlinear organic crystals, for mor information please contact the instrument responsible

General information about the Furka endstations can be found at: Furka
For questions and further information about Furka contact:  Dr Elia Razzoli

PSI User Office
Paul Scherrer Institute
building WBBC
CH-5232 Villigen PSI
Switzerland

+41 56 310 46 66
useroffice@psi.ch

Office hours:
Monday through Friday from 8:00-11:30 and 12:30-17:00
otherwise please contact us per email